Atrial synchronous ventricular pacing seems to be the best pacing mode for patients with advanced AV block and impaired LV function. The long-term follow-up of single lead VDD pacing was studied in 33 patients with impaired LV function and compared to 42 patients with normal LV function. All patients received the same VDD lead and VDDR pacemaker. The lead model with 13-cm AV spacing between the atrial and ventricular electrode was implanted in 89% of the patients. Follow-ups were 1, 3, 6, and 12 months after implantation. The percentage of atrial sensing and the P wave amplitude were determined at each follow-up. Minimal P wave amplitude at implantation was 2.0 +/- 1.4 mV in patients with impaired and 1.7 +/- 0.9 mV with normal LV function (not significant). At the 12-month follow-up, 33 patients with normal and 23 patients with depressed LV function remained paced in the VDD mode. The remaining patients died in five (impaired LV function) and seven cases (normal LV function) or their pacemakers were programmed to the VVI/VVIR pacing mode in four (impaired LV function) and three cases (normal LV function). P wave amplitude did not differ in the two groups (e.g., at month 12: impaired: 1.17 +/- 0.42 mV; normal: 1.09 +/- 0.49 mV). The atrial sensitivity was programmed in most patients to sensitive settings with no differences between the two groups (e.g., at month 12: impaired: 0.13 +/- 0.06 mV; normal: 0.13 +/- 0.05 mV). The diagnostic counters indicated nearly permanent atrial sensing (e.g., at month 12: impaired: 99.3 +/- 2.2%; normal: 99.0 +/- 1.0 mV). In conclusions, single lead VDD pacing restored AV synchronous ventricular pacing in patients with normal and with impaired LV function indicating that it could be an alternative to DDD pacemakers, but not to dual-chamber pacing.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.