This characterization of a new (103)Pd directional brachytherapy source helps to establish calibration methods that could ultimately be used in the well-established AAPM Task Group 43 formalism. Monte Carlo methods accurately predict the changes in the energy spectrum caused by the fluorescent x-rays produced in the gold shield.
Magnetic resonance imaging (MRI) offers outstanding soft tissue contrast that may reduce uncertainties in target and organ-at-risk delineation and enable online adaptive image-guided treatment. Spatial distortions resulting from non-linearities in the gradient fields and nonuniformity in the main magnetic field must be accounted for across the imaging field-of-view to prevent systematic errors during treatment delivery. This work presents a modular phantom and software application to characterize geometric distortion (GD) within the large field-of-view MRI images required for radiation therapy simulation. The modular phantom is assembled from a series of rectangular foam blocks containing high-contrast fiducial markers in a known configuration. The modular phantom design facilitates transportation of the phantom between different MR scanners and MR-guided linear accelerators and allows the phantom to be adapted to fit different sized bores or coils. The phantom was evaluated using a 1.5 T MR-guided linear accelerator (MR-Linac) and 1.5 T and 3.0 T diagnostic scanners. Performance was assessed by varying acquisition parameters to induce image distortions in a known manner. Imaging was performed using T1 and T2 weighted pulse sequences with 2D and 3D distortion correction algorithms and the receiver bandwidth (BW) varied as 250-815 Hz pixel −1 . Phantom set-up reproducibility was evaluated across independent set-ups. The software was validated by comparison with a non-modular phantom. Average geometric distortion was 0.94 ± 0.58 mm for the MR-Linac, 0.90 ± 0.53 mm
Primary S measurements of the CivaDot demonstrated good repeatability and reproducibility of the free-air chamber measurements. Measurements of the CivaDot dose distribution using the EBT3 film stack phantom and its subsequent comparison to Monte Carlo-predicted dose distributions were encouraging, given the overall uncertainties. This work will aid in the eventual realization of a clinically viable dosimetric framework for the CivaSheet based on the CivaDot dose distribution.
The differences between the SK measurements determined from the seven-distance technique and the accredited UWADCL measurement results were within the k = 2 uncertainty reported for an accredited calibration. Excellent agreement was found between the measured SK and RAKR methods used at the UWMRRC and PTB, respectively. Additionally, the sensitivity study has shown that the seven-distance algorithm accurately determines the SK of a source while having a variable chamber offset among nominal positions; the uncertainty budget for the seven-distance method does not need to be expanded at this time. It has been determined that the current standard used by the UWADCL for well chamber calibrations is valid for the Bebig model Ir2.A85-2 192Ir brachytherapy source.
The oldest form of radiation therapy, brachytherapy, has been investigated and reported in the scientific and medical literature for well over a century. Known by many names over the years, radium based, empirical practices evolved over decades to contemporary practice. This includes treatment at various dose rates using multiple radionuclides or even electrically generated photon sources. Predictions or prognostications of what may happen in the future enjoy a history that spans centuries, for example those by Nostradamus in the 1500s. In this review article, publications from several eras of past practice between the early 1900s and the late 2010s where the authors address the “future of brachytherapy” are presented, and for many of these publications, one can use the benefit of the intervening years to comment on the accuracy or the inaccuracies inherent in those publications. Finally, recently published papers are reviewed to examine current expectations for the future practice of brachytherapy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.