Patients with bilateral BHRs who have blood metal ion levels below our implant specific threshold were at low-risk of having ARMD. Cite this article: Bone Joint J 2016;98-B:1455-62.
Repeat ultrasound in MoMHA patients demonstrated that findings frequently progress in the short-term. Therefore, regular surveillance of MoMHA patients is important, with ultrasound representing an effective investigation for identifying the development and progression of lesions.
If blood metal ions are used to screen hip resurfacing patients for adverse reactions to metal debris it is recommended those with small femoral head sizes (38-44 mm) and high acetabular inclinations (>55°) are targeted. These findings require validation in other cohorts as they may not be applicable to all hip resurfacing devices given the differences in radial clearance, coverage arc, and metallurgy.
Implant design was the most important factor affecting blood metal ion concentrations. We recommend the regularity of follow-up be tailored to survival rates of various MoM THR designs rather than according to femoral head size.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.