To obtain a physiological response by a total artificial heart (TAH), while eliminating the hemodynamic abnormalities commonly observed with its use, we proposed the use of a conductance- and arterial pressure-based method (1/R control) to determine TAH cardiac output. In this study, we endeavored to make use of a variable more closely tied to central nervous system (CNS) efferents, systemic conductance, to provide the CNS with more direct control over the output of the TAH. The control equation that calculates the target cardiac output of the TAH was constructed on the basis of measurement of blood pressures and TAH flow. The 1/R control method was tested in TAH-recipient goats with an automatic method by using a microcomputer. In 1/R control animals, the typical TAH pathologies, such as mild arterial hypertension and substantial systemic venous hypertension, did not occur. Cardiac output varied according to daily activity level and exercise in a manner similar to that observed in natural heart goats. These results indicate that we have determined a control method for the TAH that avoids hemodynamic abnormalities exhibited by other TAH control systems and that exhibits physiological responses to exercise and daily activities under the conditions tested. The stability of the control and the complete lack of inappropriate excursions in cardiac output is suggestive of CNS involvement in stabilizing the system.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.