Ultralow-Power Bioimpedance IC With Intermediate Frequency Shifting Chopper

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
17
0

Year Published

2016
2016
2024
2024

Publication Types

Select...
4
2
2

Relationship

1
7

Authors

Journals

citations
Cited by 41 publications
(24 citation statements)
references
References 8 publications
0
17
0
Order By: Relevance
“…In this work, the required bandwidth and power of the IA are reduced by pre-demodulating the BioZ signal to the chopping frequency (fchop) before it enters the core IA ( Fig. 10) [16]. The main problem of a demodulator is that is tends to reduce the input impedance while increasing the noise due to additional switch on-resistance (similarly as chopping).…”
Section: B Bio-impedance (Bioz) Modementioning
confidence: 99%
See 1 more Smart Citation
“…In this work, the required bandwidth and power of the IA are reduced by pre-demodulating the BioZ signal to the chopping frequency (fchop) before it enters the core IA ( Fig. 10) [16]. The main problem of a demodulator is that is tends to reduce the input impedance while increasing the noise due to additional switch on-resistance (similarly as chopping).…”
Section: B Bio-impedance (Bioz) Modementioning
confidence: 99%
“…The main problem of a demodulator is that is tends to reduce the input impedance while increasing the noise due to additional switch on-resistance (similarly as chopping). To mitigate these effects, the IA's input chopper and the BioZ input modulator are merged into one block clocked by fMT=fBioZ-fchop [16].…”
Section: B Bio-impedance (Bioz) Modementioning
confidence: 99%
“…However, sinusoidal oscillators tend to be power hungry [3], [26][27]. On the other hand, using a square signal instead of a sinusoid causes a significant error in the impedance measurement [9], [28]. A three-level injection signal is as easy…”
Section: Bio-impedance Measurementmentioning
confidence: 99%
“…The frequency domain signal flow in an IFSC method is summarized in the right upper corner of Figure 6 [6]. The induced voltage signals (V ip<1> , V in<1> ) with the high-injection frequency are transferred to an intermediate frequency signal before amplification.…”
Section: Design Of Reconfigurable Afementioning
confidence: 99%
“…For example, an electropotential (EP) mode is used in applications including electrocardiogram (ECG), electromyogram (EMG), and electroencephalogram (EEG) [3,4,5]; impedance (IMP) mode is used for applications including body fat based on body impedance analysis [6,7], respiration rate [8,9], and galvanic skin response [10]; electrochemical (EC) mode is used in applications including blood glucose monitoring [11]; and photoelectric (PE) mode is used in applications including photoplethysmography (PPG) [12,13] and blood oxygen saturation [14,15]. Moreover, recently, as the requirements on integrated multifunctional biosignal acquisition systems are increasing, some research on multiparameter biosignal acquisition in a single system on a chip (SoC) have been reported [16,17,18].…”
Section: Introductionmentioning
confidence: 99%