The nano-Au self-assembled glassy carbon electrode for selective determination of epinephrine in the presence of ascorbic acid

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

3
48
1
2

Year Published

2008
2008
2022
2022

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 93 publications
(54 citation statements)
references
References 34 publications
3
48
1
2
Order By: Relevance
“…The linear regression equation of E pa /V ¼ 0.641 À 0.068 pH was obtained (correlation coefficient r ¼ 0.998). The slope value is close to the theoretical Nernstian value of 0.059 V indicating that the number of protons and electrons involved are equal [28,34]. Figure 3 also reveals that the i pa increases with increase in pH up to 7.0 and further increase of pH results in the decrease of anodic peak current.…”
Section: à2supporting
confidence: 70%
See 2 more Smart Citations
“…The linear regression equation of E pa /V ¼ 0.641 À 0.068 pH was obtained (correlation coefficient r ¼ 0.998). The slope value is close to the theoretical Nernstian value of 0.059 V indicating that the number of protons and electrons involved are equal [28,34]. Figure 3 also reveals that the i pa increases with increase in pH up to 7.0 and further increase of pH results in the decrease of anodic peak current.…”
Section: à2supporting
confidence: 70%
“…, which was higher than the reported D o value for EP at L-cysteine and gold nanoparticles modified gold electrodes [31,34,35]. This reveals that at the present modified electrode, the oxidation of EP occurs at faster rate and hence higher oxidation current was observed.…”
contrasting
confidence: 76%
See 1 more Smart Citation
“…The epinephrine electrooxidation at bare electrode is associated with problems of slow electron transfer rate and adsorptions of epinephrine on the electrode surfaces leading to passivation [20]. Electrochemical methods for epinephrine analysis making use of modified carbon electrodes [14 -19], carbon nanotubes [20,21] and gold nanoparticles modified electrodes [14,22,23] have been reported. The major challenge in epinephrine analysis is the elimination of interferences from ascorbic acid; ascorbic acid oxidation peak is very close to that of epinephrine which often results to their peaks overlapping.…”
Section: Introductionmentioning
confidence: 99%
“…From the point of view of medicine, it is a drug for emergency treatment in severe allergic reaction, cardiac arrest and sepsis [23]. Therefore, the quantitative determination of EP concentration in different human fluids, such as plasma and urine, is important for developing nerve physiology, pharmacological research and life science [24].…”
Section: Introductionmentioning
confidence: 99%