Abstract:Purpose:
To demonstrate by measurement that using partially shielded Gd‐153 sources for rotating‐shield brachytherapy (RSBT) is feasible. RSBT is a potentially superior alternative to conventional high‐dose‐rate brachytherapy and provides the opportunity to dramatically improve tumor dose conformity for the treatment of, for example, prostate cancer.
Methods:
A custom‐built, stainless steel encapsulated 150 mCi Gd‐153 capsule with an outer length of 12.8 mm, outer diameter of 2.10 mm, active length of 9.98 mm,… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.