volume 274, issue 1-2, P119-129 2004
DOI: 10.1016/j.ijpharm.2004.01.009
View full text
|
|
Share

Abstract: This study deals with in vitro investigation of the release of di(2-ethylhexyl)phthalate (DEHP) during hemodialysis and pump-oxygenation therapy using medical grade PVC tubing. High resolution GC-MS analysis showed that the release of DEHP was time-dependently increased by circulation of bovine blood into a major system for the hemodialysis that is used in Japan, and the amount of DEHP released into the blood had reached 7.3 mg by 4 h of circulation. No significant difference was observed in the release patter…

expand abstract