Reference level for patient dose in dental skull lateral teleradiography
Abstract: The present work describes an experiment to obtain a local reference dose value in lateral skull teleradiography for dental applications. The reference value was based on patient measurements in seven dental X-ray units, using thermoluminescent chips, and measurements on a rubber globe filled with water at another 78 installations. Dosemeters were located initially in the head and neck of a human phantom to select the most suitable locations, and on the cephalostat of the X-ray unit at two appropriate location…
Search citation statements
Paper Sections
Select...
4
1
0
0
Citation Types
0
1
0
0
Year Published
Range
2008
2019
Publication Types
Select...
5
Relationship
0
5
Authors
Journals
Cited by 5 publications
(1 citation statement)
References 12 publications
0
1
0
0
Abstract
Smart CitationsHow this paper cites the one you are viewing
“…In the past, a cephalogram with a long distance between X-ray source and mid-sagittal plane of the patient's face (3-4 m) was used. This type of machine allows more parallel X-ray beams, leading to less magnification of the images and possibly less radiation dose to the patient, when paired with sensitive image receptors (3)(4)(5)(6). Today, most of the machines on the market combine panoramic and cephalometric radiographic options within one single device.…”