Modifications of the Glaucoma Drainage Implant To Prevent Early Postoperative Hypertension and Hypotony: A Laboratory Study

Abstract: Hypotony or hypertension in the early postoperative period following implantation of a nonvalved seton such as the Molteno tube is a common problem. We conducted a laboratory investigation evaluating two modifications of the silicone drainage tube. One involved a longitudinally-oriented, pressure-sensitive slit-valve combined with an absorbable occluding ligature interposed between the valve and the episcleral plate. Experiments examining the relationship between slit length, opening pressure, and flow rate we… Show more

Search citation statements

Order By: Relevance

Paper Sections

Select...
8
1
1
0

Citation Types

0
2
0
0

Year Published

Range
1997
1997
2023
2023

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

Cited by 10 publications

(2 citation statements)

References 18 publications

0
2
0
0
Order By: Relevance
How the paper below cites the one you are viewing
“…Prior studies have indicated that aqueous humor, with its various proteins and blood products, can occlude fenestrations. [ 9 ] Aqueous humor also has a different viscosity from balanced salt solution and, as such, the actual opening pressure of an implanted BGI may be slightly different; however, we posit that the relatively small contribution from the second fenestration in most of the IOP ranges tested still holds clinically. It is important to keep in mind that the critical pressures reported in this study are likely lower than what would be seen in patients; the effect of episcleral fibrosis was not accounted for in this study.…”
Section: Discussion
mentioning
confidence: 97%