Modal analysis of HPGe detector assembly in gamma-ray spectrometers

Abstract: Abstract. The HPGe detector assembly of gamma-ray spectrometer cooled by Stirling cycle cryocooler is under consideration. Modal analysis based on the compiled dynamic model was carried out. The natural frequencies and modes shapes for HPGe detector with relative efficiency 15 % mounted by supports made from composite G-Etronax and CESTILENE HD 1000 were calculated by Solidworks simulation. The frequencies of the axial mode are in range of 200-600 Hz where electrical interferences caused by mechanical vibratio… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
3
0
3

Year Published

2018
2018
2019
2019

Publication Types

Select...
2
1

Relationship

2
1

Authors

Journals

citations
Cited by 3 publications
(6 citation statements)
references
References 9 publications
0
3
0
3
Order By: Relevance
“…Moreover, it is important to take into account the Q-factor of considered oscillatory systems, which is determined by the energy dissipation in the material and elastic supports. For example, the Q-factor of the detector assembly with fiberglass supports for the axial vibration mode does not exceed 10-15 [7]. The Q-factor of the axial resonances of the vacuum cap is much higher and reaches values of 50-70.…”
Section: The Dynamic Model Of Cryostat With Hpge Detectormentioning
confidence: 99%
See 2 more Smart Citations
“…Moreover, it is important to take into account the Q-factor of considered oscillatory systems, which is determined by the energy dissipation in the material and elastic supports. For example, the Q-factor of the detector assembly with fiberglass supports for the axial vibration mode does not exceed 10-15 [7]. The Q-factor of the axial resonances of the vacuum cap is much higher and reaches values of 50-70.…”
Section: The Dynamic Model Of Cryostat With Hpge Detectormentioning
confidence: 99%
“…The gamma-ray spectrometer "Monolith" can be equipped with HPGe crystals of various size and mass, which are determined by the required efficiency [7]. The mass of the installed crystal can vary from 0.5 to 1.8 kg or more.…”
Section: The Dynamic Model Of Cryostat With Hpge Detectormentioning
confidence: 99%
See 1 more Smart Citation
“…2. Результаты расчетов использованы при проектировании подвески ОЧГ-детектора с целью отстройки ее резонансных частот от вибрационного спектра используемого ЭМО [12,13].…”
Section: а бunclassified
“…Visaktīvāk ĪTG detektora kriostatā tiek izmantots labirinta tipa balsti [14]. Tāpat arī tiek minēti balstu pielietošanas rezultāti, izmantojot trīs elastīgos gredzenus [52]. Ir attēloti (3.12.att.)…”
Section: Dinamisks īTg Detektora Mezgla Modelisunclassified