Intra- and inter-brand accuracy of four dental radiometers

Abstract: This study measured the accuracy and precision of four commercial dental radiometers. The intra-brand accuracy was also determined. The light outputs from 14 different curing lights were measured three times using four brands of dental radiometers and the results were compared to two laboratory-grade power meters that were used as the "gold standard". To ensure proper representation, three examples of each brand of dental radiometer were used. Data collected was analyzed using ANOVA, with 95% confidence interv… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
29
0
2

Year Published

2013
2013
2020
2020

Publication Types

Select...
7
2

Relationship

0
9

Authors

Journals

citations
Cited by 38 publications
(31 citation statements)
references
References 15 publications
0
29
0
2
Order By: Relevance
“…644726). Taking into account the variability of the irradiance measurements obtained by some hand-held radiometers (14,15), in the present study the radiometer was used for monitoring the constancy of the curing unit irradiance before the polymerization of each composite material. The variations were under 10%, and value of 1090 mW/cm 2 represents the arithmetic mean of ten measurements.…”
Section: Methodsmentioning
confidence: 99%
“…644726). Taking into account the variability of the irradiance measurements obtained by some hand-held radiometers (14,15), in the present study the radiometer was used for monitoring the constancy of the curing unit irradiance before the polymerization of each composite material. The variations were under 10%, and value of 1090 mW/cm 2 represents the arithmetic mean of ten measurements.…”
Section: Methodsmentioning
confidence: 99%
“…Whilst the dental radiometers used in all but 2 of these surveys are inaccurate in absolute irradiance terms they give us an overall picture of clinical practice. 12,13 A South African survey reported a 100% satisfaction level by dentists with the performance of their light curing units even though nearly half of the units had inadequate output. 14 Prolonged irradiation time may compensate for low irradiance.…”
mentioning
confidence: 98%
“…Na presente pesquisa, a luz transmitida através de diferentes espessuras de resinas compostas foi mensurada a partir de um dispositivo simples, o radiômetro manual, o qual estava calibrado para registrar a intensidade de luz apenas entre os comprimentos de onda entre 400 e 500 nm. A aferição da intensidade de luz através de radiômetros de mão difere significativamente daquela mensurada por medidores de potência laboratorial, contudo servem como um instrumento importante para monitorar o desempenho de fontes de luz 21,22 . Ainda, é preciso considerar que esses radiômetros manuais fornecem valores relativos e não absolutos a respeito da luz emitida pelos aparelhos fotopolimerizadores 22,23 .…”
Section: Discussionunclassified
“…A aferição da intensidade de luz através de radiômetros de mão difere significativamente daquela mensurada por medidores de potência laboratorial, contudo servem como um instrumento importante para monitorar o desempenho de fontes de luz 21,22 . Ainda, é preciso considerar que esses radiômetros manuais fornecem valores relativos e não absolutos a respeito da luz emitida pelos aparelhos fotopolimerizadores 22,23 . Assim, na presente pesquisa, este dispositivo foi empregado com o objetivo de demonstrar o comportamento das duas resinas, considerando diferentes resinas e espessuras.…”
Section: Discussionunclassified