2010
DOI: 10.1016/j.nimb.2010.03.004
|View full text |Cite
|
Sign up to set email alerts
|

Induced radioactivity in a patient-specific collimator used in proton therapy

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
13
0

Year Published

2011
2011
2024
2024

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 12 publications
(14 citation statements)
references
References 16 publications
1
13
0
Order By: Relevance
“…Our results on dose equivalent from activation of brass aperture are similar to those reported by Cesana et al 48 where an ambient dose equivalent rate of about 11-12 lSv/h was obtained at the upstream surface of a brass block and 2-3 lSv/h at its downstream surface after 1 day cooling. The H * (10) value reduced to 2-3 lSv/h upstream and 1.5-2 lSv/h downstream after 1 month cooling.…”
Section: Isotopesupporting
confidence: 91%
See 1 more Smart Citation
“…Our results on dose equivalent from activation of brass aperture are similar to those reported by Cesana et al 48 where an ambient dose equivalent rate of about 11-12 lSv/h was obtained at the upstream surface of a brass block and 2-3 lSv/h at its downstream surface after 1 day cooling. The H * (10) value reduced to 2-3 lSv/h upstream and 1.5-2 lSv/h downstream after 1 month cooling.…”
Section: Isotopesupporting
confidence: 91%
“…The H * (10) value reduced to 2-3 lSv/h upstream and 1.5-2 lSv/h downstream after 1 month cooling. The brass that was used in Cesana et al 48 study was SAE 660 containing about two times more lead than C3600 used in our study. The slight difference in the H*(10) between our values and those of Cesana et al 48 could be explained by the difference in alloys used in the simulations.…”
Section: Isotopementioning
confidence: 78%
“…The report of task group 136 of the American Association of Physicists in Medicine proposed to survey activated materials routinely and dispose or recycle the patient-specific consumable components after completing the course of PBT and decay to the background has occurred. 11 Additionally, although handling of the activated material on the proton beam lines, such as the range compensator and the brass collimator, has been discussed, 12 , 13 the effect of the activation of metals inside the human body, such as dental metal, has not been clarified in PBT and is not specified in the guideline. In the present study, we revealed that the induced radionuclides with half-lives of several hours were produced from the dental metals commonly used in Japan.…”
Section: Discussionmentioning
confidence: 99%
“…Therefore, the radioactivation effects of induced radionuclides and their radiation exposure have been investigated. 12 , 13 These studies report the radioactivation of materials outside the human body. However, the radioactivation effect caused by materials inside the human body, such as dental metal, is yet to be clarified.…”
Section: Introductionmentioning
confidence: 99%
“…The activated materials are generally subject to radiation control via the emission of radiation such as beta and gamma rays [2]. Several studies have reported the radioactivation effects of patient-specific devices in PBT [3,4]. Although they researched the radioactivation of materials outside a human body, it may also occur in patients following irradiation in PBT, and several reports have focused on this phenomenon for real-time monitoring and range verification by positron emission tomography imaging [5][6][7][8][9].…”
Section: Introductionmentioning
confidence: 99%