2010
DOI: 10.1088/0031-9155/55/12/018
|View full text |Cite
|
Sign up to set email alerts
|

Improved dose-calculation accuracy in proton treatment planning using a simplified Monte Carlo method verified with three-dimensional measurements in an anthropomorphic phantom

Abstract: Treatment planning for proton tumor therapy requires a fast and accurate dose-calculation method. We have implemented a simplified Monte Carlo (SMC) method in the treatment planning system of the National Cancer Center Hospital East for the double-scattering beam delivery scheme. The SMC method takes into account the scattering effect in materials more accurately than the pencil beam algorithm by tracking individual proton paths. We confirmed that the SMC method reproduced measured dose distributions in a hete… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
42
0

Year Published

2013
2013
2017
2017

Publication Types

Select...
7

Relationship

3
4

Authors

Journals

citations
Cited by 22 publications
(42 citation statements)
references
References 21 publications
0
42
0
Order By: Relevance
“…Such a simplification serves to reduce the calculation time of the dose deposition in a voxel. Accuracy of the SMC method was confirmed by experiments 14 , 15 . The estimated mean statistical error of the total dose calculation in the planning target volume (PTV) with the optimized spot weights was 0.57% rms of the maximum dose in the PTV.…”
Section: Methodsmentioning
confidence: 82%
See 2 more Smart Citations
“…Such a simplification serves to reduce the calculation time of the dose deposition in a voxel. Accuracy of the SMC method was confirmed by experiments 14 , 15 . The estimated mean statistical error of the total dose calculation in the planning target volume (PTV) with the optimized spot weights was 0.57% rms of the maximum dose in the PTV.…”
Section: Methodsmentioning
confidence: 82%
“…A simplified Monte Carlo (SMC) method 12 , 13 , 14 , 15 has been developed to obtain dose calculation results more accurately than the PBA in target heterogeneities in a shorter calculation time. It tracks individual proton paths scattered in material and uses a measured depth‐dose distribution in water for dose calculation.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…For DMU calc,PSD and DMU calc,BSD calculations, we used the simplified Monte Carlo method (SMC) 12 , 13 . It tracks individual protons in a range compensator, an aperture collimator, and a phantom by calculating the range loss and scattering in materials using the water‐equivalent model (16) and the Highlands's equation.…”
Section: Methodsmentioning
confidence: 99%
“…However, since it takes much time to calculate the DMU with the full Monte Carlo method, reduction of calculation time is required to apply the Monte Carlo method for calculating the DMU in a patient routinely. Meanwhile, we have already developed a simplified Monte Carlo (SMC) method, with which we can quickly calculate dose distributions accurately 12 , 13 …”
Section: Introductionmentioning
confidence: 99%