volume 14, issue 1, P23-29 1980
DOI: 10.1002/jbm.820140104
View full text
|
|
Share

Abstract: Three-dimensional finite element analysis (FEA) has been used to determine the implant-bone interface characteristics of bioglass dental implants. The results of the FEA were verified by comparison with the results of mechanical testing performed on animal implant specimens. The results of the study showed that the assumption of a discontinuous change of elastic properties at the bone-implant interface was a poor assumption for the bioglass implants. Interface elastic moduli of 354.0, 155.0, and 47.0 MPa for c…

Expand abstract