Hong Xu, Chun-Feng Duan, Zhi-Feng Zhang, Jia-Yi Chen, Chun-Ze Lai, Mei Lian, Li-Juan Liu, Hua Cui

Abstract: A novel flow injection procedure was developed for the determination of p-aminophenol (PAP) based on the inhibition by PAP of the chemiluminescence from luminol-dimethylsulfoxide (DMSO)-NaOH-EDTA system. The method has merits of higher sensitivity, wider linear range, simpler procedure, and a more rapid analyzing speed. It is applicable for the determination of PAP in the range of 2.5 x 10(-10)-5.0 x 10(-8) g mL(-1) with a detection limit of 1.9 x 10(-10) g mL(-1). The relative standard deviation (RSD) for 5.0…