Evaluation of permanent I-125 prostate implants using radiography and magnetic resonance imaging

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

1
71
1

Year Published

1998
1998
2016
2016

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 142 publications
(73 citation statements)
references
References 16 publications
1
71
1
Order By: Relevance
“…63 Another problem is postoperative edema, which typically increases the prostate volume by 40 to 50% compared to the preoperative volume. 59,60,66 If the postimplant CT scan is obtained immediately after the implant is performed, the dose may be underestimated. The edema increases the distance between the seeds, as well as the volume, thereby lowering the dose rate.…”
Section: Technical Issuesmentioning
confidence: 99%
See 1 more Smart Citation
“…63 Another problem is postoperative edema, which typically increases the prostate volume by 40 to 50% compared to the preoperative volume. 59,60,66 If the postimplant CT scan is obtained immediately after the implant is performed, the dose may be underestimated. The edema increases the distance between the seeds, as well as the volume, thereby lowering the dose rate.…”
Section: Technical Issuesmentioning
confidence: 99%
“…Fusion of CT and MR images may also be a viable solution, as MR provides adequate visualization of the prostate while CT provides localization of the implanted seeds. 66,67 These numerous difficulties and technical challenges notwithstanding, the standards for seed implant quality are being defined in terms of quantitative CT-based dosimetric evaluation. Willins and Wallner 68 reported that, for CT scans obtained on the day of implantation, coverage of 80% or more of the target volume by the prescription dose is probably adequate.…”
Section: Technical Issuesmentioning
confidence: 99%
“…To make the time integration in Eq. (1) expressible in a closed analytic form for easy analysis, we have parameterized the overall dose fall-off around the radioactive source, g(r)/r 2 , by a single inverse power function of distance, ∞r −τ , such that the instantaneous dose rate at P in presence of edema became, (5) where Ḋ P (0), identical to the same quantity in Eq. (2), is thus related to the conventional dose prescription in absence of edema.…”
Section: B Edema-induced Dose Reductionmentioning
confidence: 99%
“…3). The implant quality can be evaluated after implantation with the use of CT [42]or MRI [43]. Roy et al [42]analyzed 10 patients treated with 125 I by CT scanning and demonstrated a 150–Gy coverage in 78–96% of cases.…”
Section: Brachytherapymentioning
confidence: 99%