Evaluation of Hemolysis in the VentrAssist Implantable Rotary Blood Pump

Abstract: The VentrAssist implantable rotary blood pump (IRBP) is an implantable centrifugal blood pump with a hydrodynamically suspended impeller; optimal efficiency requires small running clearances (70-300 microm). The effect of running clearance and polish on hemolysis was evaluated in vitro. Three different human blood suspensions were compared: phosphate buffered saline (PBS), plasma volume expander (Hemaccel), and whole blood. The test conditions were: blood hematocrit 30%, flow rate 5 L/min, pressure across pump… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

5
38
2
1

Year Published

2006
2006
2020
2020

Publication Types

Select...
8
2

Relationship

0
10

Authors

Journals

citations
Cited by 43 publications
(46 citation statements)
references
References 8 publications
5
38
2
1
Order By: Relevance
“…However, our results are consistent with previous studies, which indicate low haemolysis in the axial HeartMate II system in patients [21] and negligible haemolysis by the centrifugal VentrAssist device in experimental studies [22,23]. These observations Fig.…”
Section: Discussionsupporting
confidence: 94%
“…However, our results are consistent with previous studies, which indicate low haemolysis in the axial HeartMate II system in patients [21] and negligible haemolysis by the centrifugal VentrAssist device in experimental studies [22,23]. These observations Fig.…”
Section: Discussionsupporting
confidence: 94%
“…31 The reported incidence of hemolysis for the HeartMate II axial rotary pump is 0.06, 0.02, and 0.13 events per patient-year in its BTT U.S., destination therapy U.S., and European multicenter clinical trials, respectively. 33 In a study comparing hemolysis indices in CF LVAD patients 34 supported by the axial CF rotary pump (HeartMate II) and the (VentrAssist) [35][36] centrifugal CF rotary pump, patients with the centrifugal LVAD displayed normal haptoglobin levels, low plasma free hemoglobin, and moderately increased LDH. In comparison, the HeartMate II patients showed lower haptoglobin and higher free hemoglobin and LDH values.…”
Section: Hemolysismentioning
confidence: 99%
“…This condition was near the condition on which the CFD analysis was performed. Details concerning the experimental conditions can be found elsewhere (18,19). The citrate phosphate dextran (CPD) anticoagulated fresh porcine blood was used as a working fluid at 24 ° C. The normalized index of hemolysis (NIH) was calculated using the following the equation (Eq.…”
Section: Hemolysis Testmentioning
confidence: 99%