1983
DOI: 10.1055/s-2007-1022004 View full text |Buy / Rent full text
|
|

Abstract: Twenty-three months after mitral valve replacement for mitral stenosis with a St. Jude Medical heart valve prosthesis size 31, a 29-year-old man suddenly went into a profound cardiogenic shock and pulmonary edema due to escape of one of the leaflets from the prosthesis. At emergency operation the valve was replaced with another type of prosthesis. The escaped leaflet was retrieved from the abdominal aorta and was found to be intact. The cause of the escape was found to be a fracture in the pivot area. Further … Show more

Help me understand this report

This publication either has no citations yet, or we are still processing them

Set email alert for when this publication receives citations?

See others like this or search for similar articles