1998
DOI: 10.1097/00003446-199812000-00006
|View full text |Cite
|
Sign up to set email alerts
|

Effects of Background Noise on Click-Evoked Otoacoustic Emissions

Abstract: The findings of the study indicate ambient noise levels for accurate CEOAE recording should not exceed 50 to 55 dB A of noise and alternatives to the default program should be considered in non-sound-treated situations.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

3
32
1
3

Year Published

2002
2002
2021
2021

Publication Types

Select...
6
2

Relationship

1
7

Authors

Journals

citations
Cited by 49 publications
(40 citation statements)
references
References 29 publications
3
32
1
3
Order By: Relevance
“…Like other measures of auditory function such as otoacoustic emissions [Rhoades et al, 1998;Smith et al, 2001], ambient acoustic noise can be a confounding variable which may affect the accuracy of ASSR measurements. It is necessary to separate the signal (genuine ASSRs) from the noise (defined as all activities other than the ASSRs).…”
Section: Introductionmentioning
confidence: 99%
“…Like other measures of auditory function such as otoacoustic emissions [Rhoades et al, 1998;Smith et al, 2001], ambient acoustic noise can be a confounding variable which may affect the accuracy of ASSR measurements. It is necessary to separate the signal (genuine ASSRs) from the noise (defined as all activities other than the ASSRs).…”
Section: Introductionmentioning
confidence: 99%
“…The Reflwin (Costa et al 2009;Rhoades et al 1998;Shi et al 2000). In the present study, nine reference standards were established using the AABR, HFT, TEOAE and DPOAE tests or their combinations for the determination of test performance of WBA.…”
Section: Wideband Absorbancementioning
confidence: 99%
“…Ukoliko je spoljašnji šum širokopojasni nestacionarni šum onda on utiče na smanjenje SNR i reproduktivnosti u delu spektra u kome se javlja. Istraživanje Rhoadesa i saradnika (Rhoades et al, 1998) .…”
Section: Uticaj Spoljašnjih Maskirajućih šUmova Na Merenje Oaeunclassified
“…Pored pomenutih, na merenje OAE mogu uticati kijanje, pomeranje jabučice i mišića. Navedeni procesi imaju uticaja na rezultate merenja u oblasti između 500 i 1500 Hz (Rhoades et al, 1998;Whitehead et al, 1993;Jacobson, 1994;Lasky, 1998;Tlumak, 2001). …”
Section: Uticaj Bioloških šUmova Na Merenje Oaeunclassified
See 1 more Smart Citation