Effect of nebulizer type and antibiotic concentration on device performance
Abstract: We compared the performance of selected ultrasonic and jet nebulizers when aerosolizing several antibiotic formulations to determine optimum combinations for delivery of a respirable antibiotic aerosol. Three ultrasonic devices were tested: the UltraNeb 99/100ா, the UltraAIRா and the Aerosonicா. The reusable jet nebulizers were the Dura ProNebா, Pari-LLா and the Sidestreamா. The six disposable jet nebulizers were Marquest Acorn IIா, Hudson T Updraftா II, Baxter MistyNebா, Pari-LCா, Pari IS-2ா, and a disposable…
Search citation statements
Paper Sections
Select...
12
1
1
0
Citation Types
1
11
0
0
Year Published
1999
2025
Publication Types
Select...
14
Relationship
0
14
Authors
Journals
Cited by 14 publications
(12 citation statements)
References 23 publications
1
11
0
0
“…This is not the first study to illustrate the in vitro differences in nebulizer performance, but it is unique in demonstrating the impact of the V D on the efficiency of a device with regards to total nebulized drug output when the effect of evaporative losses are taken into consideration. The RF observed for the four devices is within the range reported in previous studies (14,15,19,29). Hess et al (6) reported V D data, unfortunately nebulizer specific comparisons were not possible because the data in this study were pooled for all nebulizer types for different flow and volume settings.…”
supporting
confidence: 79%
“…This is not the first study to illustrate the in vitro differences in nebulizer performance, but it is unique in demonstrating the impact of the V D on the efficiency of a device with regards to total nebulized drug output when the effect of evaporative losses are taken into consideration. The RF observed for the four devices is within the range reported in previous studies (14,15,19,29). Hess et al (6) reported V D data, unfortunately nebulizer specific comparisons were not possible because the data in this study were pooled for all nebulizer types for different flow and volume settings.…”
supporting
confidence: 79%
“…Nebulizer aerosol delivery is largely dependent on the nebulizer design and brand [55,56]. In this study, the effect of nebulizer design on improving tobramycin aerosol delivery has been studied by using different nebulizer delivery systems.…”
Section: Resultsmentioning
confidence: 99%
“…The tobramycin concentration was determined by using validated HPLC method following pre-column derivatization with fluorescein isothiocyanate (FITC) [30]. The Guideline on Bioanalytical Method Validation.…”
Section: Tobramycin Assaymentioning
confidence: 99%