Eur J Chem volume 10, issue 2, P102-107 2019 DOI: 10.5155/eurjchem.10.2.102-107.1813
Randa Bakheet Ahmed, Mohamed El-Muktar Abdelaziz, Ahmed Elsadig Mohammed Saeed
Abstract:

A simple, precise, reproducible and economical HPLC method for estimation of tinidazole has been developed. The wavelength of 317 nm is selected as λmax for tinidazole in phosphate buffer (pH = 6.8). Validation parameters were tested following International Conference on Harmonization (ICH) guideline. Tinidazole shows linearity at the selected wavelength and obeys Beer’s law in the concentration range of 3.2-40.0 µg/mL with correlation coefficient of 0.9999. Recovery studies for tinidazole were performed a…