Detection of inner tube defects in co‐axial circle and Bain breathing systems: a comparison of occlusion and Pethick tests

Abstract: SummaryThe performance of the occlusion and Pethick tests in detecting faulty inner tubes in co-axial circle and Bain systems was compared. Twelve co-axial circle and 12 Bain anaesthetic breathing systems were tested using the occlusion and the Pethick tests. For each system, three tubes were intact, and the remaining nine had a defect deliberately created in the inner tube (three proximal, three middle and three distal). The investigators were blinded to which of the tubes were defective, and to each other's … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
7
0

Year Published

2011
2011
2023
2023

Publication Types

Select...
5
2

Relationship

0
7

Authors

Journals

citations
Cited by 8 publications
(7 citation statements)
references
References 6 publications
0
7
0
Order By: Relevance
“…[89]Pethick test: This tests the low-pressure system and the integrity of the inner tube. The procedure is as follows: Occlude the patient end of the circuit.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…[89]Pethick test: This tests the low-pressure system and the integrity of the inner tube. The procedure is as follows: Occlude the patient end of the circuit.…”
Section: Discussionmentioning
confidence: 99%
“…Collapse of the reservoir bag due to the creation of venturi effect in the outer tube is an indication that the inner tube is intact. [8] This test will not detect a system in which inner tube is omitted or does not extend to the patient port or one that has holes at the patient end of the inner tube.Foex-Crampton-Smith manoeuvre: This manoeuvre assesses the gas flow line from the flowmeters of the machine to the patient end of the circuit. With an oxygen flow at 2 litres/minute, the patient end of the inner tube is occluded briefly for 2–3 seconds using the forefinger.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Fresh gas flow was then reduced to the anesthesia machine's minimum (250 mL/min). The breathing system was considered intact if pressure was maintained at 30 cmH 2 O for 5 seconds; the breathing system was deemed faulty if the pressure failed to reach 30 cmH 2 O or was not maintained for five seconds [15]. Relative humidity and temperature were measured at the endotracheal tube at 5 minutes and again at 30 minutes with a hygrometer (Drager, Serial No: 044936, model 5204644-A4).…”
Section: Methodsmentioning
confidence: 99%
“…Although coaxial systems are convenient to use, problems have been associated with a failure of the integrity of the inner lumen, resulting in re-breathing and consequently potential patient harm [3,4]. In order to ensure integrity of the inner tube specific occlusion tests have been recommended [5,6].…”
Section: Introductionmentioning
confidence: 99%