Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear

Abstract: SIGNIFICANCE As scleral lens wear becomes more common, understanding the impact of these lenses upon ocular physiology is critically important. Studies on the effect of scleral lens wear upon intraocular pressure (IOP) have used different instruments and have reported conflicting results. PURPOSE The purpose of this study was to compare assessment of IOP during scleral lens wear using pneumatonometry and transpalpebral tonometry. METHODS Twen… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
28
1

Year Published

2021
2021
2024
2024

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 15 publications
(29 citation statements)
references
References 31 publications
0
28
1
Order By: Relevance
“…The first methodological challenge is the instrument to be used, as IOP cannot be measured with the accepted gold standard technique (Goldmann applanation tonometry) with the lens in-situ. Measurements obtained following lens removal show minimal change in IOP after short term lens wear [274][275][276][277][278], although one study reported an average increase of 5.8 mmHg after 8 h of wear [279]. Only two studies have examined post-lens removal IOP following longer periods of lens wear (up to 6 months) and reported no significant variation [280].…”
Section: Intraocular Pressurementioning
confidence: 99%
“…The first methodological challenge is the instrument to be used, as IOP cannot be measured with the accepted gold standard technique (Goldmann applanation tonometry) with the lens in-situ. Measurements obtained following lens removal show minimal change in IOP after short term lens wear [274][275][276][277][278], although one study reported an average increase of 5.8 mmHg after 8 h of wear [279]. Only two studies have examined post-lens removal IOP following longer periods of lens wear (up to 6 months) and reported no significant variation [280].…”
Section: Intraocular Pressurementioning
confidence: 99%
“…12 The stable IOP observed after lens wear (median change<0.1 mm Hg increase) is also consistent with recent studies of IOP measured after nonfenestrated scleral lens (15–18 mm diameters) removal using nontrans palpebral tonometers (mean changes ranging from a 1.3 mm Hg decrease to a 1.0 mm Hg increase). 29–35…”
Section: Discussionmentioning
confidence: 99%
“…A majority of studies take two to three consecutive IOP measurements for each eye at each specific time point and average the results, but of note, such studies typically make no comment regarding the confidence interval or standard deviation readout from the instrument itself (the instrument’s assessment of the data reliability). 1 , 2 , 4 , 6 , 12 We believe reporting on this data set is essential as this is how a clinician in a typical office setting would determine whether a measurement taken individually was “reliable” by manufacturer’s standards. Each individual pressure measurement deemed “reliable” and analyzed in each cohort for each eye in the handheld tonometry group had a CI of 90% or greater (mean 94.38%) and in the pneumatonometry group had a standard deviation less than 2 mmHg (mean 0.468).…”
Section: Discussionmentioning
confidence: 99%
“…In recent years, clinicians have sought to determine whether the application and wear of a prosthetic replacement of the ocular surface ecosystem (PROSE, BostonSight, Needham MA) or scleral lens induces any change in intraocular pressure (IOP). 1–6 Prosthetic replacement of the ocular surface ecosystem (PD) is an integrated treatment that includes designing, manufacturing and fitting a customized medical ocular surface prosthetic device (PD), comprehensive application and removal training, and ongoing collaborative optimization of optometric and ophthalmic care. The PD is manufactured using a gas-permeable plastic material (typically fluorosilicone acrylate) which is customized to each eye using a web-based design software (FitConnect, BostonSight, Needham MA) and mechanical lathe.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation