Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients

Abstract: The ICare tonometer can be useful in a routine clinical setting. The IOP readings are quite in accordance with those obtained by GAT. The measurements seemed to be influenced by CCT variations, and thus pachymetry should always be taken into consideration.

Search citation statements

Order By: Relevance

Paper Sections

Select...
124
57
24
3

Citation Types

19
140
4
16

Year Published

Range
2000
2000
2026
2026

Publication Types

Select...
173
8
5

Relationship

3
183

Authors

Journals

Cited by 186 publications

(179 citation statements)

References 32 publications

19
140
4
16
Order By: Relevance
How the paper below cites the one you are viewing
“…This was similar to the 100 µm increase in CCT per 1 mmHg difference in IOP readings reported by Pakrou et al [ 2 ]. However, our study reported a lesser influence of CCT on difference in IOP measurements as compared to Brusini et al [ 1 ] who reported a CCT change of 10 μm per 0.7 mmHg deviation in iCare, which may be due to their lower sample size compared to ours.…”