Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients
Abstract: The ICare tonometer can be useful in a routine clinical setting. The IOP readings are quite in accordance with those obtained by GAT. The measurements seemed to be influenced by CCT variations, and thus pachymetry should always be taken into consideration.
Search citation statements
Paper Sections
Select...
124
57
24
3
Citation Types
19
140
4
16
Year Published
Range
2000
2026
Publication Types
Select...
173
8
5
Relationship
3
183
Authors
Journals
Cited by 186 publications
(179 citation statements)
References 32 publications
19
140
4
16
Smart CitationsHow the paper below cites the one you are viewing
“…This was similar to the 100 µm increase in CCT per 1 mmHg difference in IOP readings reported by Pakrou et al [ 2 ]. However, our study reported a lesser influence of CCT on difference in IOP measurements as compared to Brusini et al [ 1 ] who reported a CCT change of 10 μm per 0.7 mmHg deviation in iCare, which may be due to their lower sample size compared to ours.…”
Section: Discussion
contrasting
confidence: 96%
Abstract