volume 9, issue 2, P151-167 1975
DOI: 10.1002/jbm.820090205
View full text
|
|
Share

Abstract: The measured tensile and fatigue strengths of wrought and heat-treated cobalt chromium-molybdenum-carbon alloys such as HS21 were found to be more than twice those of as cast alloys of the same composition. The resistance of wrought HS21 to crevice corrosion at metal-Teflon contacts in isotonic salt solution at 37 degrees C was found to be considerably greater than that of cast HS21, wrought HS25, wrought MP35, or 316L stainless steel which are currently used for surgical implants. The increased crevice corros…