Biomechanical Evaluation of the Fatigue Performance, the Taper Corrosion and the Metal Ion Release of a Dual Taper Hip Prosthesis under Physiological Environmental Conditions

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
13
0
3

Year Published

2018
2018
2022
2022

Publication Types

Select...
7
1
1

Relationship

0
9

Authors

Journals

citations
Cited by 20 publications
(16 citation statements)
references
References 53 publications
0
13
0
3
Order By: Relevance
“…The reason for this is, that a not firmly fixed modular neck/stem taper connection is of high risk for mechanically assisted crevice corrosion. A loosened neck/stem taper connection can be originated during index surgery by an insufficient assembling force or can be caused by macroscopic visible material loss at the medial neck taper interface due to corrosion resulting in a toggling of the modular CoCr neck relatively to the Ti alloy stem [79].…”
Section: Decision Making Model For Asymptomatic Dual Taper Stems In Cmentioning
confidence: 99%
See 1 more Smart Citation
“…The reason for this is, that a not firmly fixed modular neck/stem taper connection is of high risk for mechanically assisted crevice corrosion. A loosened neck/stem taper connection can be originated during index surgery by an insufficient assembling force or can be caused by macroscopic visible material loss at the medial neck taper interface due to corrosion resulting in a toggling of the modular CoCr neck relatively to the Ti alloy stem [79].…”
Section: Decision Making Model For Asymptomatic Dual Taper Stems In Cmentioning
confidence: 99%
“…In addition, the restoration of the centre of rotation and related ligament balancing during trial head reduction possibly requires a higher offset or longer neck length [4,7,80,81]. Increasing the offset or neck length to adapt for a different cup orientation or sufficient soft tissue balancing may create a more demanding biomechanical loading situation [7,16,32,[82][83][84] for the so far asymptomatic dual taper neck adapter, possibly triggering MACC [17,32,79]. Therefore an indication may be given for revision of the modular stem.…”
Section: Decision Making Model For Asymptomatic Dual Taper Stems In Cmentioning
confidence: 99%
“…In addition, a second electrolyte containing 10 g/l FeCl 3 in Ringer's solution (FeCl 3 ) (B. Braun AG, Melsungen, Germany) was used for electrochemical tests of polished, deep-rolled and microblasted Ti discs. The FeCl 3 -containing electrolyte has a low pH of 1.8 and a higher concentration of Cl − ions and was used as corrosion accelerating environment [26,27]. Electrochemical testing was done at 37 • C and by using aerated electrolytes.…”
Section: Electrochemical Testingmentioning
confidence: 99%
“…Thus far, loosening studies indicate the importance of the dimensions of hip endoprosthesis [3], applied materials [4] and body characteristics of patients [5]. Another significant cause of failures, fatigue fractures, may happen due to inadequate endoprosthesis [6] and tribological processes [7]. In addition, inadequate dimensions and materials choice may make unpleasant changes for patients in terms of the hip joint squeaking [8].…”
Section: Introductionmentioning
confidence: 99%