An evaluation of the mini-Wright peak flow meter.

Abstract: The mini-Wright peak flow meter (MPFM) has been evaluated, and the results obtained from it show a strong positive correlation (r= 0-970) with the Wright's peak flow meter (PFM). MPFM measurements, however, were biased to be about 38 1/min higher than PFM measurements (95% confidence limits 31[0 1/min to 450 1/min). Between instrument variation was found (F-ratio 3 67 with 9 and 81 degrees of freedom: P<0 001). In practice this did not appreciably affect individual measurements greatly as 95% confidence limits… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

2
24
0
2

Year Published

1981
1981
2014
2014

Publication Types

Select...
9
1

Relationship

0
10

Authors

Journals

citations
Cited by 61 publications
(28 citation statements)
references
References 6 publications
2
24
0
2
Order By: Relevance
“…7 The statistical methods suggested by Bland and Altman4 were not used in these studies, but biases estimated at 3%1 and 38 I/min7 were thought to be clinically acceptable and similar to variations between different standard Wright meters. Small variations between instruments were noted with the mini Wright meters but were not regarded as clinically important.…”
Section: Discussionmentioning
confidence: 99%
“…7 The statistical methods suggested by Bland and Altman4 were not used in these studies, but biases estimated at 3%1 and 38 I/min7 were thought to be clinically acceptable and similar to variations between different standard Wright meters. Small variations between instruments were noted with the mini Wright meters but were not regarded as clinically important.…”
Section: Discussionmentioning
confidence: 99%
“…A 20,21 . Posteriormente, esse valor foi expresso em porcentagem do valor previsto para a altura e sexo do paciente, utilizando-se a equação de Torres et al 22 Após a avaliação inicial, a terapêutica e os controles foram instituídos, seguindo as recomendações internacionais estabelecidas em 1995 2 .…”
Section: Métodosunclassified
“…The PEF was measured using the mini-Wright peak flow meter. 15,16 The parameters were assessed before and 20 minutes after the inhalation of 400μg of Salbutamol using a metered dose inhaler (MDI) and a spacer device. For spirometry, a minimum of three and a maximum of eight spirometry maneuvers were done according to ERS/ATS criteria.…”
Section: Lung Functionmentioning
confidence: 99%