2017
DOI: 10.1016/j.arth.2016.08.025
|View full text |Cite
|
Sign up to set email alerts
|

Accelerometer-Based, Portable Navigation (KneeAlign) vs Conventional Instrumentation for Total Knee Arthroplasty: A Prospective Randomized Comparative Trial

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
64
0

Year Published

2017
2017
2024
2024

Publication Types

Select...
4
3

Relationship

0
7

Authors

Journals

citations
Cited by 44 publications
(64 citation statements)
references
References 21 publications
0
64
0
Order By: Relevance
“…1). [8][9][10][15][16][17][18][19][20][21][22][23][24] Among the included studies, six were RCTs, 8,10,17,19,21,23 one was a prospective study, 16 and six were retrospective studies. 9,15,18,20,22,24 The present meta-analysis analyzed 1,454 patients with 1,551 TKAs (mean sample size, 119.31 TKAs per study).…”
Section: Search Resultsmentioning
confidence: 99%
See 2 more Smart Citations
“…1). [8][9][10][15][16][17][18][19][20][21][22][23][24] Among the included studies, six were RCTs, 8,10,17,19,21,23 one was a prospective study, 16 and six were retrospective studies. 9,15,18,20,22,24 The present meta-analysis analyzed 1,454 patients with 1,551 TKAs (mean sample size, 119.31 TKAs per study).…”
Section: Search Resultsmentioning
confidence: 99%
“…All six RCTs were claimed as being randomized; five of those provided adequate generation of random sequences and appropriate concealment of allocation. 8,10,17,19,23 None of those studies completely reported blinding of the participant and personnel. Only the studies of Ikawa et al, 17 Kinney et al, 19 and Thiengwittayaporn et al 8 clearly expounded the blinding of outcome assessment.…”
Section: Quality Assessmentmentioning
confidence: 99%
See 1 more Smart Citation
“…10 Several studies have shown that despite the surgeons' experience, conventional guides can still be unreliable and it has a limited degree of accuracy. [11][12][13] Errors can happen as this technique is dependent on surgeons' judgment, fixation of instruments, knowledge of knee kinematics and hand eye coordination. 13,14 Mahaluxmivala et al analyzed 673 TKA's and found that 25% of cases had Varus and/or Valgus of ±3° regardless of the surgeons' experience.…”
Section: Introductionmentioning
confidence: 99%
“…12,18 Recently navigation systems have been developed using accelerometer and gyroscopes in attempt to combine the accuracy associated with CAS and convenience of conventional systems which does not require the use of large computer systems for registration and alignment results to provide the surgeon intra operatively. 5,6,9,11,19 The iAssist (Zimmer Inc, Warsaw, IN) system uses four pods which are attached to the surgical instruments and intraoperatively using accelerometer and gyroscope provides the precise alignment and position in relation to anatomic landmarks. Within these pods are inertial gyroscopes that exchange information using a secure local wireless channel, and information is displayed to the surgeon intraoperatively.…”
Section: Introductionmentioning
confidence: 99%