volume 29, issue 5, P609-618 1996
DOI: 10.1016/0021-9290(95)00107-7
View full text
|
|
Share
M.J. King, J. Corden, T. David, J. Fisher

Abstract: The flow through a bileaflet mechanical heart valve during the first half of systole was predicted using computational fluid dynamics (CFD). A three-dimensional model of the geometry of the ventricle, valve, sinus and aorta was developed. Flow through the valve was assumed to be Newtonian and laminar. The peak systolic Reynolds number was 1500 based on the aortic radius and the mean aortic velocity. Flow visualisation and laser Doppler anemometry (LDA) experiments were performed and the results were compared t…