Journal of Chromatography A 2012 DOI: 10.1016/j.chroma.2011.11.036 View full text
|
|
Share
James F. Griffith, William L. Winniford, Kefu Sun, Rob Edam, Jim C. Luong

Abstract: A simple and reliable differential flow modulator has been demonstrated which reverses the flow during the flush step. The modulator is constructed with commercially available capillary flow technology tees which simplifies the apparatus and permits wide range of column dimensions to be used because the modulator volume is adjustable. Using a reverse flush arrangement the tailing of the peak at the base (baseline rise between modulations) is reduced 10-20 fold as compared to forward flush modulation. This is m…