2008
DOI: 10.1016/j.apradiso.2007.10.020
|View full text |Cite
|
Sign up to set email alerts
|

A new Certified Reference Material for radionuclides in Irish sea sediment (IAEA-385)

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

4
9
0

Year Published

2010
2010
2020
2020

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 54 publications
(19 citation statements)
references
References 10 publications
4
9
0
Order By: Relevance
“…If the ICP-MS/MS is equipped with an APEX-Q high-efficiency sample introduction system, the detection limit could be improved to 0.1 fg mL -1 for 239 Pu and 0.02 fg mL -1 for 240 Pu. The analytical results of reference materials show that the measured activity concentration of 239 Pu and 240 Pu and are 1.81±0.07 mBq g -1 and 1.17±0.10 mBq g -1 in the IAEA-385 and 0.37±0.01 mBq g -1 and 0.23±0.01 mBq g -1 in the IAEA-412, respectively (Table 2), which are in a good agreement with the certified values of 1.79±0.28 mBq g -1 and 1.15±0.14 mBq g -1 of IAEA-385 [37] and 0.36±0.016 mBq g -1 and 0.24±0.016 mBq g -1 of IAEA-412 [38], respectively. All these features prove that the developed method for determination of low level 239 Pu and 240 Pu in the environmental samples is reliable and accurate.…”
Section: Resultssupporting
confidence: 86%
“…If the ICP-MS/MS is equipped with an APEX-Q high-efficiency sample introduction system, the detection limit could be improved to 0.1 fg mL -1 for 239 Pu and 0.02 fg mL -1 for 240 Pu. The analytical results of reference materials show that the measured activity concentration of 239 Pu and 240 Pu and are 1.81±0.07 mBq g -1 and 1.17±0.10 mBq g -1 in the IAEA-385 and 0.37±0.01 mBq g -1 and 0.23±0.01 mBq g -1 in the IAEA-412, respectively (Table 2), which are in a good agreement with the certified values of 1.79±0.28 mBq g -1 and 1.15±0.14 mBq g -1 of IAEA-385 [37] and 0.36±0.016 mBq g -1 and 0.24±0.016 mBq g -1 of IAEA-412 [38], respectively. All these features prove that the developed method for determination of low level 239 Pu and 240 Pu in the environmental samples is reliable and accurate.…”
Section: Resultssupporting
confidence: 86%
“…Although the 240 Pu/ 239 Pu atom ratios are not certified, the ratios obtained for the reference samples (Table 4) agree with other reported values using various analytical techniques such as ICP-MS, accelerator mass spectrometry, thermal ionisation mass spectrometry and high-resolution ␣-spectrometry [6,19,[27][28][29]32,[35][36][37][38][39][40][41][42][43][44][45][46]. The precisions (n = 10) of the measurements ranged from 1.7% to 8.6% depending on the total Pu concentrations in the samples.…”
Section: Validation Of Analytical Methodssupporting
confidence: 84%
“…The 240 Pu/ 239 Pu atom ratio (0.188 ± 0.013) in the IAEA-385 sample agrees well (within the 95% confidence level) with the information value of 0.17 ± 0.06 (k = 2) given by IAEA [35].…”
Section: Validation Of Analytical Methodssupporting
confidence: 78%
“…; Pham et al. ) helped to maintain good quality of results. A Bratislava‐Gran Sasso interlaboratory comparison carried out using the sample #19 showed results within 1 sigma standard deviation.…”
Section: Samples and Methodsmentioning
confidence: 99%