2006
DOI: 10.1590/s0100-39842006000300011
|View full text |Cite
|
Sign up to set email alerts
|

Calibração das câmaras de ionização para feixes de tomografia computadorizada no Brasil: a realidade atual

Abstract: OBJECTIVE: The aim of this study was to establish a calibration methodology specific for pencil ionization chambers used in computed tomography dosimetric procedures, in compliance with the most recent recommendations. The study was developed at the Calibration Laboratory of the Instituto de Pesquisas Energéticas e Nucleares. MATERIALS E METHODS: An industrial X-ray equipment, several types of ionization chambers, a mobile collimator (diaphragm type), and several high purity aluminum filters were utilized in t… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
2
0
5

Year Published

2008
2008
2013
2013

Publication Types

Select...
2
2

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(7 citation statements)
references
References 7 publications
0
2
0
5
Order By: Relevance
“…For the calibration, radiochromic film strips were inserted into the chamber cap (16) and positioned inside the phantom bores, reproducing exactly the position of the 10X5-3CT 100 mm pencil-type ionization chamber (Radcal Corporation) calibrated according to the established procedures (21) . The obtained results were compared with the purpose of obtaining the respective calibration coefficients.…”
Section: Methodsmentioning
confidence: 99%
“…For the calibration, radiochromic film strips were inserted into the chamber cap (16) and positioned inside the phantom bores, reproducing exactly the position of the 10X5-3CT 100 mm pencil-type ionization chamber (Radcal Corporation) calibrated according to the established procedures (21) . The obtained results were compared with the purpose of obtaining the respective calibration coefficients.…”
Section: Methodsmentioning
confidence: 99%
“…Plane-parallel ionization chambers for low-energy X-radiation and beta-radiation (13) , high-energy electrons (14)(15)(16) , computed tomography (17,18) , radioprotection (19) , extrapolation chambers for X-radiation and beta-radiation of dermatologic and ophthalmologic applications utilized in brachytherapy (20,21) have been constructed by Instituto de Pesquisas Energéticas e Nucleares.…”
mentioning
confidence: 99%
“…para realizar calibração de instrumentos com radiação gama e X, e calibração de monitores de contaminação alfa e/ou beta [18] . O procedimento de calibração deve ser meticuloso e reprodutível, para tanto o LCI utiliza qualidades recomendadas pela norma 61267 da Comissão Eletrotécnica Internacional (IEC) seguindo a Qualidade de Radiação de Referência (RQR) [75,76,77] . As tabelas 5 e 6 apresentam as características dos RQRs utilizados no Laboratório.…”
Section: Visedunclassified
“…intensidade da radiação X. [75,76] Figura 7: C.I. monitora posicionada na saida do feixe do equipamento emissor de raios X, no LCI.…”
Section: Visedunclassified