Background-Radiofrequency ablation inside pulmonary vein (PV) ostia can cause PV stenosis. A novel alternative method of ablation is irreversible electroporation, but the long-term response of PVs to electroporation ablation is unknown. Methods and Results-In ten 6-month-old pigs (60-75 kg), the response of PVs to circular electroporation and radiofrequency ablation was compared. Ten consecutive, nonarcing, electroporation applications of 200 J were delivered 5 to 10 mm inside 1 of the 2 main PVs, using a custom-deflectable, 18-mm circular decapolar catheter. Inside the other PV, circular radiofrequency ablation was performed using 30 W radiofrequency applications via an irrigated 4-mm ablation catheter. PV angiograms were made before ablation, immediately after ablation, and after 3-month survival. PV diameters and heart size were measured. With electroporation ablation, PV ostial diameter decreased 11±10% directly after ablation, but had increased 19±11% after 3 months. With radiofrequency ablation, PV ostial diameter decreased 23±15% directly after ablation and remained 7±17% smaller after 3 months compared with preablation diameter despite a 21±7% increase in heart size during aging from 6 to 9 months. Conclusions-In this porcine model, multiple circumferential 200-J electroporation applications inside the PV ostia do not affect PV diameter at 3-month follow-up. Radiofrequency ablation inside PV ostia causes considerable PV stenosis directly after ablation, which persists after 3 months. (Circ Arrhythm Electrophysiol. 2014;7:734-738.)
Background
Irreversible electroporation (IRE) using direct current (DC) is an effective method for the ablation of cardiac tissue. A major drawback of the use of DC-IRE, however, are two problems: requirement of general anesthesia due to severe muscle contractions and the formation of bubbles containing gaseous products from electrolysis. The use of high-frequency alternating current (HF-IRE) is expected to solve both problems, because HF-IRE produces little to no muscle spasms and does not cause electrolysis.
Methods
In the present study, we introduce a novel asymmetric, high-frequency (aHF) waveform for HF-IRE and present the results of a first, small, animal study to test its efficacy.
Results
The data of the experiments suggest that the aHF waveform creates significantly deeper lesions than a symmetric HF waveform of the same energy and frequency (
p
= 0.003).
Conclusion
We therefore conclude that the use of the aHF enhances the feasibility of the HF-IRE method.
Five-year freedom of atrial tachyarrhythmia could be achieved by PVAI as primary ablation strategy in 58.3% of patients. Additional substrate modification only moderately increased overall success.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.