A model based upon steady-state diffusion theory which describes the radial dependence of diffuse reflectance of light from tissues is developed. This model incorporates a photon dipole source in order to satisfy the tissue boundary conditions and is suitable for either refractive index matched or mismatched surfaces. The predictions of the model were compared with Monte Carlo simulations as well as experimental measurements made with tissue simulating phantoms. The model describes the reflectance data accurately to radial distances as small as 0.5 mm when compared to Monte Carlo simulations and agrees with experimental measurements to distances as small as 1 mm. A nonlinear least-squares fitting procedure has been used to determine the tissue optical properties from the radial reflectance data in both phantoms and tissues in vivo. The optical properties derived for the phantoms are within 5%-10% of those determined by other established techniques. The in vivo values are also consistent with those reported by other investigators.
Percutaneous treatment of portal venous stenoses is effective and long lasting in children with reduced-size hepatic transplants. In patients with elastic or recurrent lesions, portal venous stents have excellent long-term primary patency despite continued patient growth. Successful, percutaneous transhepatic venoplasty eliminates the need for surgical revision, portacaval shunting, or repeat transplantation.
Using spatially resolved, steady state diffuse reflectometry, a directional dependence was found in the propagation of visible and near infrared light through human skin in vivo. The skin's reduced scattering coefficient mu(s)' varies by up to a factor of two between different directions of propagation at the same position. This anisotropy is believed to be caused by the preferential orientation of collagen fibres in the dermis, as described by Langer's skin tension lines. Monte Carlo simulations that examine the effect of partial collagen fibre orientation support this hypothesis. The observation has consequences for non-invasive diagnostic methods relying on skin optical properties, and it could be used non-invasively to determine the direction of lines of cleavage in order to minimize scars due to surgical incisions.
Diabetics would benefit greatly from a device capable of providing continuous noninvasive monitoring of their blood glucose levels. The optical scattering coefficient of tissue depends on the concentration of glucose in the extracellular fluid. A feasibility study was performed to evaluate the sensitivity of the tissue reduced scattering coefficient in response to step changes in the blood glucose levels of diabetic volunteers. Estimates of the scattering coefficient were based on measurements of the diffuse reflectance on the skin at distances of 1-10 mm from a point source. A correlation was observed between step changes in blood glucose concentration and tissue reduced scattering coefficient in 30 out of 41 subjects measured.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.