Objectives
To compare intraoperative 3D fluoroscopy with a ceiling-mounted flat panel detector in plate osteosynthesis of distal radius fractures (AO/OTA 2R3C1.2) with volar locking plate systems to conventional 2D fluoroscopy for detection of insufficient fracture reduction, plate misplacement and protruding screws.
Methods
Using a common volar approach on 12 cadaver forearms, total intraarticular distal radius fractures were induced, manually reduced and internally fixated with a 2.4 distal radius locking compression plate. 2D (anterior-posterior and lateral) and 3D (rotational) fluoroscopic images were taken as well as computed tomographies. Fluoroscopic images, Cone Beam CT (CBCT), 360° rotating sequences (so called “Movies”) and CT scans were co-evaluated by a specialist orthopedic surgeon and a specialist radiologist regarding quality of fracture reduction, position of plate, position of the three distal locking screws and position of the three diaphyseal screws. In reference to gold standard CT, sensitivity and specifity were analyzed.
Results
“Movie” showed highest sensitivity for detection of insufficient fracture reduction (88%). Sensitivity for detection of incorrect position of plate was 100% for CBCT and 90% for “Movie.” For intraarticular position of screws, 2D fluoroscopy and CBCT showed highest sensitivity and specifity (100 and 91%, respectively). Regarding detection of only marginal intraarticular position of screws, sensitivity and specifity of 2D fluoroscopy reached 100% (CBCT: 100 and 83%). “Movie” showed highest sensitivity for detection of overlapping position of screws (100%). When it comes to specifity, CBCT achieved 100%. Regarding detection of only marginal overlapping position of screws, 2D fluoroscopy and “Movie” showed highest sensitivity (100%). CBCT achieved highest specifity (100%).
Conclusion
As for assessment of quality of fracture reduction and detection of incorrect position of plate as well as overlapping position of the three diaphyseal screws CBCT and “Movie” are comparable to CT – especially when combined. Particularly sensitivity is high compared to standard 2D fluoroscopy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.