Purpose
Radiation exposure is affected by C-arm fluoroscopy device positioning during percutaneous renal puncture. Our aim was to compare the exposure of surgeon's lens, hand and chest with a fluoroscopy protocol replicated in different C-arm positions.
Materials and Methods
A standardized fluoroscopy protocol was created using water-equivalent solid phantoms to replicate a surgeon and patient. 111 mGy radiation (360 s) was applied in standard fluoroscopy mode (91 kVp, 2.7 mA/mAs). Dosimeters were placed on lens, chest and hand of surgeon and patient phantom models. 7 different C-arm positions were created: 0°, mediolateral (ML) +90°, ML -90°, ML +30°, ML -15°, craniocaudal (CC) +30°, CC +15°. Measurements were evaluated separately for different positions.
Results
The highest radiation exposure was measured on patient dosimeter (2.97 mSv). The highest exposure on surgeon was recorded on finger dosimeter in all C-arm positions; highest dose was recorded in ML +90° position (2.88 mSv). In finger dosimeters, lowest exposure was recorded in 0° position (0.51 mSv). The lowest exposure of all positions was measured in chest dosimeter in ML -90° position (0.24 mSv).
Conclusions
In positions where X-ray generator of the C-arm was facing towards the surgeon, radiation exposure measured in all dosimeters was higher compared to positions where the generator was facing away. The hand radiation exposure in all positions was higher than chest and lens. Special care must be taken to avoid facing the X-ray generator tube and hands should be as well-protected as chest and eyes with special protective gear.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.