and the Alzheimer's Disease Neuroimaging InitiativePurpose: To create a standardized, MRI-compatible, lifesized phantom of the brain ventricles to evaluate ventricle segmentation methods using T 1 -weighted MRI. An objective phantom is needed to test the many different segmentation programs currently used to measure ventricle volumes in patients with Alzheimer's disease.
Materials and Methods:A ventricle model was constructed from polycarbonate using a digital mesh of the ventricles created from the 3 Tesla (T) MRI of a subject with Alzheimer's disease. The ventricle was placed in a brain mold and surrounded with material composed of 2% agar in water, 0.01% NaCl and 0.0375 mM gadopentetate dimeglumine to match the signal intensity properties of brain tissue in 3T T 1 -weighted MRI. The 3T T 1 -weighted images of the phantom were acquired and ventricle segmentation software was used to measure ventricle volume.
Results:The images acquired of the phantom successfully replicated in vivo signal intensity differences between the ventricle and surrounding tissue in T 1 -weighted images and were robust to segmentation. The ventricle volume was quantified to 99% accuracy at 1-mm voxel size.
Conclusion:The phantom represents a simple, realistic and objective method to test the accuracy of lateral ventricle segmentation methods and we project it can be extended to other anatomical structures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.