A simple method for the off-line coupling of laser desorption mass spectrometry (LDMS) and capillary zone electrophoresis (CZE) is described. Representative mass spectra of subpicomole quantities of proteins isolated from CZE are presented and discussed. The current detection limit for bovine alpha-lactalbumin is 100 fmols injected onto the CZE column. Horse heart myoglobin was demonstrated to be stable in CHES/KCl, a CZE buffer, for at least 1 month, suggesting that some isolates can be safely stored for long time periods prior to LDMS analysis. Protein stability in 0.1% aqueous trifluoroacetic acid (TFA), a common solvent for LDMS, must also be considered. In the special case of porcine pepsinogen, significant (greater than 50%) degradation was observed within 5 min in TFA. In favorable cases, mass measurement accuracies of +/- 0.02% were obtained for protein isolates. Factors limiting mass measurement accuracy are presented. Finally, the possibility of identifying protein isolates, by combining N-terminal sequencing, molecular mass measurements, and selective peptide "mapping" procedures, is discussed.
EXPERIMENTAL Capillary electrophoresis. The apparatus, Fig. l(a), consisted of a Hippotronics (model R50B, Brewster, NY) 0-50kV high voltage power supply. Platinum
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.