The VentrAssist implantable rotary blood pump (IRBP) is an implantable centrifugal blood pump with a hydrodynamically suspended impeller; optimal efficiency requires small running clearances (70-300 microm). The effect of running clearance and polish on hemolysis was evaluated in vitro. Three different human blood suspensions were compared: phosphate buffered saline (PBS), plasma volume expander (Hemaccel), and whole blood. The test conditions were: blood hematocrit 30%, flow rate 5 L/min, pressure across pump 100 mm Hg, 6 h flow period, and 37 degrees C. Normalized Index of Hemolysis (NIH) for the Biomedicus BP-80, used as a control, was: 0.0040 +/- 0.0023 (n = 9; x +/- SD) and 0.00014 +/- 0.00009 (n = 5) for pooled blood suspensions in PBS and Hemaccel respectively, and 0.00053 +/- 0.0002 (n = 3) in whole blood. Hemolysis was reduced by improved surface finish and unaffected by running clearance. NIH for the VentrAssist IRBP with 0.2 microm Ra surface finish was 0.000167 +/- 0.00007 (n = 4) g/100 L in whole human blood, demonstrating minimal hemolysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.