Magnesium
alloys have gained significant attention as degradable
implant materials, but the fast and localized corrosion behavior leading
to hydrogen gas evolution and alkaline poisoning limits their clinical
application. In this research, the possibility of controlling the
fast degradation rate of an experimental Mg–Si–Sr alloy
by applying hybrid biopolymer chitosan (CS)–gelatin (G)–bioactive
glass (BG) coatings was investigated. Electrophoretic deposition using
alternating current fields (AC-EPD) was employed for surface coating
and the influence of suspension parameters (biopolymer type and concentration,
BG particle size), and key AC-EPD parameters (voltage amplitude, frequency,
and time) on the coating quality were investigated. Stable suspensions
of positively charged biopolymer/BG particles deposited on the Mg
alloy coupled as a cathode during the high-amplitude peak. Furthermore,
coating homogeneity improved with increasing peak-to-peak-voltage
and the hybrid nature of the coatings was confirmed by scanning electron
microscopy and Fourier transform infrared spectroscopy. Corrosion
studies revealed a significantly decreased corrosion rate down to
0.08 mm/year for the Mg–Si–Sr alloy incorporating CS–G–BG
b AC-EPD coating.
The wear of a novel temporomandibular joint (TMJ) prosthesis was evaluated in an animal model. The prosthesis consisted of an additively manufactured titanium alloy (Ti6Al4V) mandibular condyle and glenoid fossa created through selective laser melting, with a machined vitamin E‐enriched ultra‐high molecular weight polyethylene (UHMWPE) surface attached to the fossa. Thirteen TMJ prosthesis were implanted in sheep, six of which had condylar heads coated with HadSat® diamond‐like carbon (H‐DLC). Euthanasia took place after 288 days, equaling 22 years of human mastication. Linear and volumetric wear analysis of the fossa was performed by optical scanning. The condylar head surfaces were assessed by scanning electron and confocal laser microscopy. The average linear UHMWPE wear, when combined with the coated condyle, was 0.67 ± 0.28 mm (range: 0.34–1.15 mm), not significantly differing (p = .3765, t‐test) from the non‐coated combination average (0.88 ± 0.41 mm; range: 0.28–1.48 mm). The respective mean volumetric wear volumes were 25.29 ± 11.43 mm3 and 45.85 ± 22.01 mm3, not significantly differing (p = .1448, t‐test). Analysis of the coated condylar surface produced a mean Ra of 0.12 ± 0.04 μm and Sa of 0.69 ± 0.07 μm. The non‐coated condylar surface measured a mean Ra of 0.28 ± 0.17 μm and Sa of 2.40 ± 2.08 μm. Both Sa (p = .0083, Mann–Whitney U test) and Ra (p = .0182, Mann–Whitney U test), differed significantly. The prosthesis exhibits acceptable wear resistance and addition of the H‐DLC‐coating significantly improved long‐term condylar surface smoothness.
Orthopedic device-related infections remain a serious challenge to treat. Central to these infections are bacterial biofilms that form on the orthopedic implant itself. These biofilms shield the bacteria from the host immune system and most common antibiotic drugs, which renders them essentially antibiotic-tolerant. There is an urgent clinical need for novel strategies to prevent these serious infections that do not involve conventional antibiotics. Recently, a novel antibiofilm coating for titanium surfaces was developed based on 5-(4-bromophenyl)-N-cyclopentyl-1-octyl-1H-imidazol-2-amine as an active biofilm inhibitor. In the current study we present an optimized coating protocol that allowed for a 5-fold higher load of this active compound, whilst shortening the manufacturing process. When applied to titanium disks, the newly optimized coating was resilient to the most common sterilization procedures and it induced a 1 log reduction in biofilm cells of a clinical Staphylococcus aureus isolate (JAR060131) in vitro, without affecting the planktonic phase. Moreover, the antibiofilm effect of the coating in combination with the antibiotic cefuroxime was higher than cefuroxime treatment alone. Furthermore, the coating was successfully applied to a human-scale fracture fixation device resulting in a loading that was comparable to the titanium disk model. Finally, an in vivo biocompatibility and healing study in a rabbit osteotomy model indicated that these coated implants did not negatively affect fracture healing or osteointegration. These findings put our technology one step closer to clinical trials, confirming its potential in fighting orthopedic infections without compromising healing.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.