The use of electronic portal imaging devices (EPIDs) is a promising method for the dosimetric verification of external beam, megavoltage radiation therapy-both pretreatment and in vivo. In this study, a previously developed EPID back-projection algorithm was modified for IMRT techniques and applied to an amorphous silicon EPID. By using this back-projection algorithm, two-dimensional dose distributions inside a phantom or patient are reconstructed from portal images. The model requires the primary dose component at the position of the EPID. A parametrized description of the lateral scatter within the imager was obtained from measurements with an ionization chamber in a miniphantom. In addition to point dose measurements on the central axis of square fields of different size, we also used dose profiles of those fields as reference input data for our model. This yielded a better description of the lateral scatter within the EPID, which resulted in a higher accuracy in the back-projected, two-dimensional dose distributions. The accuracy of our approach was tested for pretreatment verification of a five-field IMRT plan for the treatment of prostate cancer. Each field had between six and eight segments and was evaluated by comparing the back-projected, two-dimensional EPID dose distribution with a film measurement inside a homogeneous slab phantom. For this purpose, the y-evaluation method was used with a dose-difference criterion of 2% of dose maximum and a distance-to-agreement criterion of 2 mm. Excellent agreement was found between EPID and film measurements for each field, both in the central part of the beam and in the penumbra and low-dose regions. It can be concluded that our modified algorithm is able to accurately predict the dose in the midplane of a homogeneous slab phantom. For pretreatment IMRT plan verification, EPID dosimetry is a reliable and potentially fast tool to check the absolute dose in two dimensions inside a phantom for individual IMRT fields. Film measurements inside a phantom can therefore be replaced by EPID measurements.
The purpose of this study was to investigate the dose-response characteristics, including ghosting effects, of an amorphous silicon-based electronic portal imaging device (a-Si EPID) under clinical conditions. EPID measurements were performed using one prototype and two commercial a-Si detectors on two linear accelerators: one with 4 and 6 MV and the other with 8 and 18 MV x-ray beams. First, the EPID signal and ionization chamber measurements in a mini-phantom were compared to determine the amount of buildup required for EPID dosimetry. Subsequently, EPID signal characteristics were studied as a function of dose per pulse, pulse repetition frequency (PRF) and total dose, as well as the effects of ghosting. There was an over-response of the EPID signal compared to the ionization chamber of up to 18%, with no additional buildup layer over an air gap range of 10 to 60 cm. The addition of a 2.5 mm thick copper plate sufficiently reduced this over-response to within 1% at clinically relevant patient-detector air gaps (> 40 cm). The response of the EPIDs varied by up to 8% over a large range of dose per pulse values, PRF values and number of monitor units. The EPID response showed an under-response at shorter beam times due to ghosting effects, which depended on the number of exposure frames for a fixed frame acquisition rate. With an appropriate build-up layer and corrections for dose per pulse, PRF and ghosting, the variation in the a-Si EPID response can be reduced to well within +/- 1%.
The potential for detrimental incidents and the ever increasing complexity of patient treatments emphasize the need for accurate dosimetric verification in radiotherapy. For this reason, all curative treatments are verified, either pretreatment or in vivo, by electronic portal imaging device (EPID) dosimetry in the Radiation Oncology Department of The Netherlands Cancer Institute-Antoni van Leeuwenhoek hospital, Amsterdam, The Netherlands. Since the clinical introduction of the method in January 2005 until August 2009, treatment plans of 4337 patients have been verified. Among these plans, 17 serious errors were detected that led to intervention. Due to their origin, nine of these errors would not have been detected with pretreatment verification. The method is illustrated in detail by the case of a plan transfer error detected in a 5 x 5 Gy intensity-modulated radiotherapy (IMRT) rectum treatment. The EPID reconstructed dose at the isocenter was 6.3% below the planned value. Investigation of the plan transfer chain revealed that due to a network transfer error, the plan was corrupted. 3D analysis of the acquired EPID data revealed serious underdosage of the planning target volume: On average 11.6%, locally up to 20%. This report shows the importance of in vivo (EPID) dosimetry for all treatment plans as well as the ability of the method to assess the dosimetric impact of deviations found.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.