Purpose
Pulsed-field ablation (PFA) is a new energy source to achieve pulmonary vein isolation (PVI) by targeted electroporation of cardiomyocytes. Experimental and controlled clinical trial data suggest good efficacy of PFA-based PVI. We aimed to assess efficacy, safety and follow-up of PFA-based PVI in an early adopter routine care setting.
Methods
Consecutive patients with symptomatic paroxysmal or persistent atrial fibrillation (AF) underwent PVI using the Farawave® PFA ablation catheter in conjunction with three-dimensional mapping at two German high-volume ablation centers. PVI was achieved by applying 8 PFA applications in each PV.
Results
A total of 138 patients undergoing a first PVI (67 ± 12 years, 66% male, 62% persistent AF) were treated. PVI was achieved in all patients by deploying 4563 applications in 546 PVs (8.4 ± 1.0/PV). Disappearance of PV signals after the first application was demonstrated in 544/546 PVs (99.6%). More than eight PFA applications were performed in 29/546 PVs (6%) following adapted catheter positioning or due to reconnection as assessed during remapping. Mean procedure time was 78 ± 22 min including pre- and post PVI high-density voltage mapping. PFA catheter LA dwell-time was 23 ± 9 min. Total fluoroscopy time and dose area product were 16 ± 7 min and 505 [275;747] cGy*cm2. One pericardial tamponade (0.7%), one transient ST-elevation (0.7%) and three groin complications (2.2%) occurred. 1-year follow-up showed freedom of arrhythmia in 90% in patients with paroxysmal AF (n = 47) and 60% in patients with persistent AF (n = 82, p = 0.015).
Conclusions
PFA-based PVI is acutely highly effective and associated with a beneficial safety and low recurrence rate.
Graphical abstract
Background: A novel multielectrode radiofrequency balloon (RFB) catheter has been released for pulmonary vein isolation (PVI).
Methods and Results:In this observational study consecutive patients with drug-refractory paroxysmal or persistent atrial fibrillation (AF) undergoing first-time PVI were enrolled in 2 high-volume ablation centers. All procedures were conducted in conjunction with a 3D-mapping system. Clinical, procedural and ablation parameters were systematically analyzed. 105 patients (58% male; 52% paroxysmal AF, 68±11.3 years mean age, left atrial volume index 38.6±14.8 mL/m 2 ) were included. 241/412 (58.5%) PVs were successfully isolated with a single shot (SS), with a time-to-isolation of 11.6±8 s. Total number of radiofrequency applications was 892 (mean 2.2/PV), resulting in successful isolation of 408/412 (99%) PVs at the end of the procedure. Mean electrodes' impedance drop was significantly higher in the SS-PVI compared with non-SS applications (21.5±6.6 vs. 18.6±6.5 Ohm). Concordantly, higher temperature rise was observed in the SS vs. non-SS applications (10.9±4.9°C vs. 9.6±4.7°C).
Conclusions:In this multicenter real-world study, mean impedance drop and temperature rise were associated with successful SS-PVI applying the novel RFB catheter. These parameters may help to guide efficient usage of the new RF balloon.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.