Purpose
To assess differences between frame‐based and cone beam computed tomography (CBCT)‐defined stereotactic space and to identify predictors of the observed findings.
Methods and materials
Differences between frame‐based and CBCT‐defined stereotactic space after image co‐registration were reviewed for 529 patients. Treatment planning system reported the information about the shifts in X, Y, and Z coordinates of the center of the stereotactic space (i.e., coordinate X = 100 mm, Y = 100 mm, and Z = 100 mm) defined by the frame, and the maximum shot displacement (MSD) in mm. We collected the potential predictors of the differences. In total, 19 factors were investigated. We used multiple linear regression to evaluate associations with the increased differences.
Results
Rotational and translational shifts greater than 1° and 1 mm, respectively, were observed in 2.6% of patients. At the same time, a decrease in tumor coverage of more than 5% was detected in 8.3% of cases. It was revealed that the higher fiducial errors (both mean and maximum), the greater weight of the patient, and the lower Karnofsky Performance Scale were predictors of increased rotational, translational shifts, and the MSD.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.