The optimal surface finish for a cemented THA stem is still debated. We hypothesized surface finish would influence survival of Kerboull cemented hip arthroplasties and a matte finish would have lower survival. We reviewed survival of 433 total hip arthroplasties in 395 patients: 284 consecutive patients (310 hips) were enrolled in a prospective, randomized study of polished (165 hips) or matte finish stems (145 hips) and compared to a historical series of satin stems (123 hips) in 111 patients. The satin and matte finish implants had similar geometry but the polished was quadrangular rather than oval. Finish roughnesses were: polished (radius, 0.04 lm), satin (radius, 0.9 lm), and matte (radius, 1.7 lm). The mean age of the patients at the time of the index arthroplasty was 63.6 years. The survival rate at 13 years, using radiographic loosening as the end point, was 97.3% ± 2.6% for polished stems, 97.1% ± 2.1% for satin stems, and 78.9% ± 5.8% for matte stems. The data suggest survival of Kerboull stems was higher with a polished or satin surface finish than with a matte finish.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.