A study using live rabbits, with subsequent analysis of excised corneas stained with trypan blue/alizarin red, showed that torsional ultrasound led to less corneal endothelial cell damage than classical longitudinal ultrasound.
A study using live rabbits, with subsequent analysis of excised corneas stained with trypan blue/alizarin red, showing simulated swirling lenticular debris may contribute significantly to corneal endothelium damage during phacoemulsification.
Background: To evaluate the safety, efficacy, and predictability of laser-assisted in situ keratomileusis (LASIK) enhancement after primary LASIK and compare to Food and Drug Administration (FDA) criteria. Methods: Patients who underwent LASIK enhancement after primary LASIK between 2002 and 2019 were compared to those who underwent LASIK without retreatment. Patient demographics, preoperative characteristics, visual outcomes, and postoperative complications were compared between groups. Epithelial ingrowth (EI) development was stratified based on duration between primary and secondary procedures. Results: We compared 901 eyes with LASIK enhancement to 1127 eyes without retreatment. Age, sex, surgical eye, sphere, cylinder, and spherical equivalent (SE) were significantly different between groups (p < 0.05). At 12 months post-enhancement, 86% of the eyes had an uncorrected distance visual acuity of 20/20 or better and 93% of eyes were within ±0.50 D of the target. Development of EI (6.1%) demonstrated an odds ratio of 16.3 in the long-term compared to the short-term (95% CI: 5.9 to 45.18; p < 0.0001). Conclusions: Older age at primary LASIK, female sex, right eye, and larger sphere, cylinder and SE were risk factors for enhancement. Risk of EI significantly increased when duration between primary and enhancement procedures exceeded five years. LASIK enhancements produce favorable outcomes and meet FDA benchmarks for safety, efficacy, and predictability.
To compare the protection of the corneal endothelium provided by a cohesive and a dispersive ophthalmic viscosurgical device (OVD) against damage from torsional ultrasound and simulated lenticular debris during phacoemulsification.
Purpose
To evaluate the effect of each of the tip sizes available for the Oertli CataRhex3
®
phacoemulsification machine on efficiency.
Methods
Porcine lenses were fixed in formalin for 2 hours, then cut into 3.0 mm cubes. We studied three Oertli tips, all of which had a 30-degree bevel: easyTip 2.2 mm (20G), easyTip 2.8 mm (19G), and CO-MICS (21G). For the 19G and 20G tips, vacuum was set at 600 mmHg, irrigation rate at 50 mL/min, continuous power 70%, and bottle height 85 cm. For the 21G tip, vacuum was set at 450 mmHg; irrigation and power settings were identical to those used for the easyTip tips. We measured time to removal and chatter events to determine efficiency.
Results
Results from 20 trials for each tip showed that the larger the gauge size, the more quickly lens fragments were removed. Chatter events demonstrated an increasing trend with smaller tip gauge. The 19G tip used an average time to fragment removal of 2.8 seconds; the 20G, 3.2 seconds; and the 21G, 4.6 seconds. Increasing tip diameter from 21G to 20G decreased emulsification time by 33% (P = 0.02). Increasing the diameter from 21G to 19G further decreased time to emulsification by 42% (P = 0.003). The 21G tip had a mean 1.4 events/cube; 20G, 0.35 events; and 19G, 0.1 events. Differences in mean chatter events for each tip were each statistically significant.
Conclusion
These data suggest that when evaluated by chatter events and emulsification time, the 2.8 mm (19G) easyTip proves to have greatest efficiency.
Purpose
To determine the actual vacuum pressure generated by the Oertli CataRhex 3
®
(Oertli), using an external measuring system.
Methods
The effective vacuum pressure created by the Oertli was measured with a pressure device that was continuous with the vacuum tubing system while closed to the external environment. Measurements were taken with the machine set to 300 and 500 mmHg at flow rates of 20, 35, and 50 mL/min and at bottle heights of 60, 80, and 100 cm. Pressures were recorded after the foot pedal was depressed to vacuum setting (second position), and the pressure was allowed to stabilize. Subsequently, it was compared to the pressure value displayed by the machine.
Results
Externally measured vacuum pressure was on average 13.02% greater (39.05 mmHg) than displayed vacuum pressure at 300 mmHg (
P
< 0.005) and 8.60% greater (42.98 mmHg) than displayed vacuum at 500 mmHg (
P
< 0.005). The average difference between displayed and measured pressure increased with increasing bottle heights.
Conclusion
On average, the vacuum pressure generated in the Oertli was found to be significantly higher than the machine’s reading when the machine was set at 300 mmHg and 500 mmHg. Adjusting vacuum had variable effects on the measured versus displayed pressure readings.
Thygeson superficial punctate keratitis (TSPK) is clinically characterized by exacerbations and remissions of gray-white opacities within the corneal epithelium, most often bilateral but may be asymmetric. Symptoms typically include photophobia, tearing, blurring, and eye irritation. Although disease progression and prognosis are well described, the exact cause is unknown. Hypotheses exist implicating virus-mediated immunity as the cause of TSPK following cases of viral keratitis; however, several polymerase chain reaction studies refute the infectious process concurrently with symptomatic TSPK. This is further supported by the consistent lack of response to antiviral and antibacterial treatment. A subset of dendritic cells known as Langerhans cells (LC) found within the corneal epithelium has been positively correlated with exacerbations of TSPK. Langerhans cells proliferate to protect and mitigate the cornea's inflammatory response, but the inflammatory triggers and relapses associated with TSPK are not well understood. Several topical drugs exist to treat inflammation related to TSPK; however, drug delivery is a major barrier to treatment because of the tear film and epithelial barrier. Drug-eluting contact lenses that target intermediates of inflammation could serve as a more effective treatment modality because of the increased bioavailability of the drugs. This review is an in-depth survey of the literature regarding the relationship between the origin and pathophysiology of LC and TSPK at the immunologic level. We also discuss potential pharmacotherapeutic interventions for TSPK prevention and treatment.
scite is a Brooklyn-based startup that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.