Discomfort related to cuff inflation may bias 24 h ambulatory blood pressure (BP) measurements, especially during night-time. We accessed the impact of cuff inflations by comparing 24 h BP recorded with a cuff-less tonometric wrist device and an upper-arm oscillometric cuff device. Fifty-three participants with type 2 diabetes were assigned randomly to four 24-h BP recordings with a cuff (TM2430: visit 1 or 2, and 4) and a tonometric device (BPro: visit 1 or 2, 3, and 4). The mean 24 h systolic BP was significantly higher when measured with the cuff versus the tonometric device (141.6±14.6 vs. 128.3±14.6 mmHg, P≤0.01), as was nocturnal BP (6.7±5.3 vs. 10.3±7.6%, P=0.002). In conclusion, nocturnal BP decline was higher when measured with the cuff device, suggesting that cuff inflations did not increase night-time BP. Further evaluation of the tonometric device using the updated European Society of Hypertension International Protocol revision 2010 is recommended before applying it in daily clinical practice.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.