The FRANCE TAVI registry provided reassuring data regarding trends in TAVR performance in an all-comers population on a national scale. Nonetheless, given that TAVR indications are likely to expand to patients at lower surgical risk, concerns remain regarding potentially life-threatening complications and pacemaker implantation. (Registry of Aortic Valve Bioprostheses Established by Catheter [FRANCE TAVI]; NCT01777828).
In patients with de novo atherosclerotic lesions of the CFA, the perioperative morbidity and mortality rate was significantly lower among patients who underwent endovascular therapy by stenting compared with surgery, whereas clinical, morphological, and hemodynamic outcomes were comparable at mid-term. (Traitement des Lésions Athéromateuses de l'Artère Fémorale Commune par Technique Endovasculaire Versus Chirurgie Ouverte [Endovascular Versus Open Repair of the Common Femoral Artery] [TECCO]; NCT01353651).
Objectives
The aim of this study was to determine the accuracy of DoseMap™ software as compared to gafchromic film in real clinical practices.
Background
The radiation exposure from cardiovascular procedures could expose patients to potential risk of cancer and/or skin injury. New tools like Dosemap software were developed to estimate the patient skin dose in the cardiac catheterization laboratory. However, little data are available to validate this estimation of patient radiation skin dose.
Methods
This is a prospective cross‐sectional study comparing the peak skin dose (PSD) measured by gafchromic film used as reference with an estimated PSD given by Dosemap software, in patients with BMI > 30 kg/m2 undergoing elective diagnostic and/or therapeutic interventional cardiology procedures, from April 2016 to December 2016, at the Brest University Hospital Centre, France.
Results
After four exclusions of patients for poor quality of gafchromic films, 90 patients were included, with 58 angiography (67.4%), 14 percutaneous interventions (16.3%), and 14 both (16.3%). The median PSDDosemap and PSDFilm were similar with 157 mGy [IQR: 99; 273] versus 158 mGy [IQR: 101; 295] (P = 0.65), respectively, with an excellent correlation (r = 0.95). The comparison between cumulative air kerma and PSDFilm was different 366 mGy [IQR: 246; 575] versus 158 mGy [IQR: 101; 295] (P < 0.01) with moderate correlation (r = 0.79). No correlation was found between the dose area product and PSDFilm (r = 0.51).
Conclusion
DoseMap is an effective and valid method as compared to gafchromic films to estimate PSDs during interventional cardiologic procedures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.