Objective:
The aim of this study was to prospectively test the performance and potential for clinical integration of software that automatically calculates the right-to-left ventricular (RV/LV) diameter ratio from computed tomography pulmonary angiography images.
Methods:
Using 115 computed tomography pulmonary angiography images that were positive for acute pulmonary embolism, we prospectively evaluated RV/LV ratio measurements that were obtained as follows: (1) completely manual measurement (reference standard), (2) completely automated measurement using the software, and (3 and 4) using a customized software interface that allowed 2 independent radiologists to manually adjust the automatically positioned calipers.
Results:
Automated measurements underestimated (P < 0.001) the reference standard (1.09 [0.25] vs1.03 [0.35]). With manual correction of the automatically positioned calipers, the mean ratio became closer to the reference standard (1.06 [0.29] by read 1 and 1.07 [0.30] by read 2), and the correlation improved (r = 0.675 to 0.872 and 0.887). The mean time required for manual adjustment (37 [20] seconds) was significantly less than the time required to perform measurements entirely manually (100 [23] seconds).
Conclusions:
Automated CT RV/LV diameter ratio software shows promise for integration into the clinical workflow for patients with acute pulmonary embolism.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.