Background
Fingerstick point-of-care and dried-blood-spot (DBS) HCV RNA testing increases testing uptake and linkage to care. This systematic review evaluated the diagnostic accuracy of point-of-care testing and DBS to detect HCV RNA.
Methods
Bibliographic databases and conference presentations were searched for eligible studies. Meta-analysis was used to pool estimates.
Results
Of 359 articles identified, 43 studies were eligible and included. When comparing the Xpert HCV Viral Load Fingerstick assay to venous blood samples (7 studies with 987 samples), the sensitivity and specificity for HCV RNA detection was 99% (95% CI:98%-100%) and 99% (95% CI:96%-100%) and for HCV RNA quantification was 100% (95% CI:93%-100%) and 100% (95% CI:94%-100%). The proportion of invalid results following Xpert HCV Viral Load Fingerstick testing was 6% (95% CI:3%-11%). When comparing DBS to venous blood samples (28 studies with 3988 samples), the sensitivity and specificity for HCV RNA detection was 97% (95% CI:95%-98%) and 100% (95% CI:98%-100%) and for HCV RNA quantification was 98% (95% CI:96%-99%) and 100% (95% CI:95%-100%).
Conclusion
Excellent diagnostic accuracy was observed across assays for detection of HCV RNA from fingerstick and DBS samples. The proportion of invalid results following Xpert HCV Viral Load Fingerstick testing highlights the importance of operator training and quality assurance programs.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.